What sample volume and concentration do I need for the Pulsoid?

Sample volume: load 65 µL into each side of the fluid cell (the sample side and the buffer side).

Concentration: the Pulsoid's measurable range is 10⁹ to 10¹² particles/mL. For best results, dilute your sample in Measurement Electrolyte (ME) to a target of 2.5 × 10⁹ to 2.5 × 10¹¹ particles/mL.

The best dilution depends on your sample and chip, so check the particle rate during your recording. Aim for 200–2,500 particles/min.

Electrolyte: use ME prepared with the NPS Reagent Kit. Don't use general laboratory PBS, as it may contain particles that interfere with your measurement.

 


 