How do I prepare my sample for the Pulsoid?

Dilute your sample in Measurement Electrolyte to get a good particle rate.

Dilute your sample in Izon's Measurement Electrolyte (ME) to a target of 2.5 × 10⁹ to 2.5 × 10¹¹ particles/mL. 

  1. Dilute your sample in ME to the target concentration.
  2. Optional: filter your sample through a 0.1 µm filter to reduce chip blocking. FILTERING MAY CHANGE YOUR SAMPLE'S COMPOSITION.
  3. Load 65 µL into the sample side of the fluid cell.
  4. Check the particle rate. Aim for 200–2,500 particles/min. If it's outside this range, adjust your dilution.

Don't know your sample's concentration? Prepare a few dilutions, for example 1:100, 1:10 and 1:5.

Entering the dilution in PCS: use a number (e.g. 1000) or E notation (e.g. 1E3). "1:1000" and "1000x" are not valid.

Check compatibility: your sample will touch PMMA, silicon and nitrile. Make sure your sample is compatible with these materials.

 

 

 