How do I prepare my sample for the Pulsoid?
Dilute your sample in Measurement Electrolyte to get a good particle rate.
Dilute your sample in Izon's Measurement Electrolyte (ME) to a target of 2.5 × 10⁹ to 2.5 × 10¹¹ particles/mL.
- Dilute your sample in ME to the target concentration.
- Optional: filter your sample through a 0.1 µm filter to reduce chip blocking. FILTERING MAY CHANGE YOUR SAMPLE'S COMPOSITION.
- Load 65 µL into the sample side of the fluid cell.
- Check the particle rate. Aim for 200–2,500 particles/min. If it's outside this range, adjust your dilution.
Don't know your sample's concentration? Prepare a few dilutions, for example 1:100, 1:10 and 1:5.
Entering the dilution in PCS: use a number (e.g. 1000) or E notation (e.g. 1E3). "1:1000" and "1000x" are not valid.
Check compatibility: your sample will touch PMMA, silicon and nitrile. Make sure your sample is compatible with these materials.