Why isn't the pressure holding or responding?

Power cycle the Pulsoid to reset the pressure system.

If the pressure won't stay at the set point, the pressure system may be leaking. Power cycle the Pulsoid.

If nothing happens when you set the pressure, power cycle the Pulsoid. Then check the Device Status panel. Pressure System should show Calibrating for about 70 seconds, then Active.

Raise a support ticket if:

  • Pressure System doesn't change to Calibrating
  • you can't hear the pressure system while it is calibrating
  • calibration takes longer than expected
  • the pressure still won't hold after power cycling

Include the serial number from the back of the instrument.

 

DO NOT OPEN THE LID WHILE PRESSURE IS APPLIED. SET THE PRESSURE TO 0 PA FIRST.

 