How do I power cycle the Pulsoid?

Restart the instrument to fix connection and pressure system problems.

Power cycling restarts the Pulsoid. Try it first if PCS can't find the instrument, the pressure isn't holding or responding, or a firmware update fails.

  1. Press and hold the power button (on top of the instrument, back right corner) until the halo lights turn off.
  2. Wait 10 seconds.
  3. Press and hold the power button until the halo lights turn back on.
  4. In PCS, check the Device Status panel. Instrument should show Connected. Pressure System shows Calibrating for about 70 seconds, then Active. When Lid shows Closed, the Pulsoid is ready to use.

If a firmware update fails, power cycling returns the Pulsoid to the previous firmware version.