Why is there regular noise in my baseline?
Symmetrical or regularly spaced noise is usually electrical interference.
The RMS noise should be below 100 pA at a sample rate of 250 ksps. If you see regular noise in the baseline:
- Remove the fluid cell and put it back on.
- Make sure the lid is closed. It shields the chip from electrical interference.
- Move sources of interference, such as mobile phones, away from the Pulsoid.
- Check for sources of vibration, such as centrifuges, on the same bench.
If the noise continues, raise a support ticket. Include the serial number from the back of the instrument.