Why is there regular noise in my baseline?

Symmetrical or regularly spaced noise is usually electrical interference.

The RMS noise should be below 100 pA at a sample rate of 250 ksps. If you see regular noise in the baseline:

  1. Remove the fluid cell and put it back on.
  2. Make sure the lid is closed. It shields the chip from electrical interference.
  3. Move sources of interference, such as mobile phones, away from the Pulsoid.
  4. Check for sources of vibration, such as centrifuges, on the same bench.

If the noise continues, raise a support ticket. Include the serial number from the back of the instrument.

 

 

 