Why is my Measurement Electrolyte showing as a custom electrolyte?
IDS2 doesn't recognise the electrolyte in your recording as Izon's Measurement Electrolyte (ME).
For size and zeta measurements, IDS2 fills in the chip's zeta potential automatically, but only if your recording uses ME exactly as listed in PCS. Otherwise, IDS2 treats it as a custom electrolyte.
If you used ME, you can fix this in one of two ways:
- Edit the recording in IDS2 to update the electrolyte, or
- Enter a Nanopore Zeta value of −11 mV when you create the calibrated measurement.
THE −11 MV VALUE ONLY APPLIES TO IZON'S ME. Don't use it for other electrolytes.
To prevent this: keep ME selected (the default) in the Electrolyte field in PCS. Only use Add New Electrolyte for other electrolytes.