Why is my baseline current railed, or much higher or lower than expected?

Check the chip and the fluid cell.

A wet chip gives a stable corrected baseline current of 16–40 nA. The exact value depends on the chip. If your current is railed, or stable but much higher or lower than this, the cause is usually a broken chip or a short circuit in the fluid cell.

  1. Remove the fluid cell. Check the contacts and pogo pins for signs of corrosion.
  2. Take the fluid cell apart. Check the O-rings seal fully and no liquid is seeping out.
  3. Reassemble the fluid cell with a new chip.
  4. If the problem continues with the new chip, raise a support ticket. Include the serial number from the back of the instrument.

 

 