Why does the AFC show 'No Column' when my qEV column is attached?

Check that your column is RFID-tagged and fitted correctly.

The AFC will only work with RFID-tagged qEV columns. Before a column has been inserted, or if a column without an RFID tag has been inserted, the lighting will be white and the home screen will display No Column.

Do not attempt to use UN-TAGGED or NON-IZON COLUMNS with the AFC.


  1. Check that you have fitted the column mount that matches the type of column you are using.
  2. Check that the Izon logo on the column is facing away from the AFC tower.
  3. Turn the column within the column mount until the type of qEV column is displayed on the touchscreen.

Ensure that the connection between the column and the adaptor is secure to avoid any leakage.

Still showing No Column?

  1. Check that your AFC has the latest firmware. Press About in the top right of the home screen to see your firmware version. To update, download the latest AFC Updater: support.izon.com/what-is-the-latest-version-of-the-afc-updater. 
    After updating, don’t forget to calibrate the scale before use (Settings > Scale Calibration).
  2. If you have another AFC, try the same column on it.

 

Good to know: The AFC instrument records the number of times that each qEV column has been used. A warning indicator will be displayed when the recommended number of uses (5) has been exceeded, and a countdown from 5 to 0 uses will begin. The AFC will display Invalid column when the column has been used a total of 10 times.

 

If the problem continues, raise a support ticket.

Please provide the serial number of the AFC (on the underside of the AFC) and the AFC firmware version.