Why do I see blockades with blank electrolyte?
There are still particles in the fluid cell. Rinse until the blockade rate is below 10 particles/min.
Blockades in clean Measurement Electrolyte (ME) mean particles are left over. Before you load the next sample, the residual blockade rate must be below 10 particles/min.
Common causes: the previous sample wasn't fully removed, the ME is contaminated with particles, or an O-ring is contaminated.
- Remove the fluid cell. Pipette out ALL the fluid from both sides. Tap the fluid cell gently on a padded surface to collect any droplets stuck at the edges, then remove them.
- Refill both sides with fresh ME. Remove any bubbles and make sure the funnels are dry.
- Put the fluid cell back and check the blockade rate. Repeat steps 1–3 until it is below 10 particles/min.
- Still too high? Make up a new ME stock, using particle-free consumables.
- Still too high? Rinse the O-rings with 70% ethanol and let them dry. Replace them if needed (spare O-rings are supplied).
Tip: Run your calibration particles before your sample. This lowers the risk of leftover sample affecting your calibration.