Why do I see blockades with blank electrolyte?

There are still particles in the fluid cell. Rinse until the blockade rate is below 10 particles/min.

Blockades in clean Measurement Electrolyte (ME) mean particles are left over. Before you load the next sample, the residual blockade rate must be below 10 particles/min.

Common causes: the previous sample wasn't fully removed, the ME is contaminated with particles, or an O-ring is contaminated.

  1. Remove the fluid cell. Pipette out ALL the fluid from both sides. Tap the fluid cell gently on a padded surface to collect any droplets stuck at the edges, then remove them.
  2. Refill both sides with fresh ME. Remove any bubbles and make sure the funnels are dry.
  3. Put the fluid cell back and check the blockade rate. Repeat steps 1–3 until it is below 10 particles/min.
  4. Still too high? Make up a new ME stock, using particle-free consumables.
  5. Still too high? Rinse the O-rings with 70% ethanol and let them dry. Replace them if needed (spare O-rings are supplied).

 

Tip: Run your calibration particles before your sample. This lowers the risk of leftover sample affecting your calibration.