How do I import and check recordings in IDS2?

Check your data before you calibrate, so you know which recordings to repeat.

Create or open a project

  1. Open IDS2 and click the Open / Create Project folder icon.
  2. Select your Pulsoid output folder, or create a new folder (right-click > New > Folder).
  3. Click Select Folder. The project takes the folder's name. You can only have one project per folder.

Import recordings
4. Click Import Recordings under Data Processing.
5. Go to your Pulsoid data folder, select the .idf2 file and click Open.

Check your data
6. Your recordings open in the Blockade Count vs Blockade Start Time plot. Untick recordings so that only one measurement group is shown.
7. Check that the lines are straight, don't overlap, and that higher pressures and larger voltages give a higher particle rate (a steeper line).
8. Repeat for each measurement group. If lines overlap or are in the wrong order, repeat those recordings.

 

Tip: Check your data as you go, so you can repeat recordings while your fluid cell is still set up.

 