How do I fill in the Recording Settings?
All fields must be filled in before you can start a recording.
Fill in the Recording Settings panel in PCS for each sample or calibration recording.
- Investigation Name: the name of the folder your recordings are saved in.
- Chip ID: the chip pack number, plus a letter for each chip (e.g. P5T-00018 A).
- Recording Type: choose Sample or Calibration. The ID field below changes to match.
- Sample ID / Calibration ID: added to the recording file name. Maximum 18 characters.
- Dilution: a number (e.g. 1000) or E notation (e.g. 1E3). "1:1000" and "1000x" are not valid.
- Electrolyte: Measurement Electrolyte (ME) is the default. Zeta analysis depends on ME. For another electrolyte, choose Add New Electrolyte at the bottom of the list.
- Mean Particle Size: the size of your calibration particles (calibration recordings only).
- Raw Concentration: the calibration particle concentration before dilution (calibration recordings only).
UPDATE THE FIELDS EVERY TIME YOU SWITCH TO A NEW SAMPLE.
Recording won't start?
- A red outline means a field is empty, contains invalid characters, or ends with a space.
- Don't use these characters: \ / : * ? " < > |
- Click Start Recording to see a list of the fields that need fixing.
- A recording can't start while the Device Status panel shows the Lid as Open.