How do I calibrate a measurement in IDS2?
Turn your sample and calibration recordings into size, concentration or zeta potential results.
IDS2 uses your calibration recordings to work out your sample's properties. Import and check your recordings first.
- Click 'Calibrate Measurement' in the Data Processing menu bar.
- Choose size and concentration, or size and zeta.
- Enter a name for the calibrated measurement.
- Select the recording(s) for ONE sample and its matching calibration recording(s).
- For size and zeta only:
- Enter the zeta potential of your calibration particles in your electrolyte. For Izon calibration particles in ME, it is printed on the vial.
- Enter the zeta potential of the chip (Nanopore Zeta). You can skip this if you used Izon's Measurement Electrolyte (ME) and selected it in PCS.
- If you used ME but IDS2 shows it as a custom electrolyte, edit the recording to update the electrolyte, or enter a Nanopore Zeta value of −11 mV. This value only applies to ME.
- Click 'Check Selection'. Orange warnings let you continue. Red warnings must be fixed first.
- Click 'Confirm'. The calibrated measurement appears under Sample Data.
- Select it to see an overview of the results. For full details, click 'Display Summary Statistics' or 'Display Blockade Summary Data' under Data Analysis.
Repeat for each sample. After your first calibration, the calibration recordings are saved as a measurement group, so you can select the whole group next time.